Software for medical devices: what to demand from your developer

EHRs and scheduling aren't medical devices; diagnostic software is. What ANVISA and the FDA ask, what IEC 62304 requires and why a SaMD-trained vendor matters.

Software for medical devices: what to demand from your developer
Índice do artigo
faltam 14 min de leitura

Software for healthcare and software for medical devices are different products, even when the code looks the same. In the first, the client buys a feature. In the second, they buy the feature and the proof of how it was made, because that proof goes into their process with ANVISA (Brazil's health regulator), the FDA or the European notified body: how each function was specified, what risk it introduces, how it was verified and where that is written down.

That difference changes what a medical equipment manufacturer, a healthtech or a clinic network should demand from whoever develops for them. And it also changes the profile of who is on the other side of the table. That is why X-Apps invests in training its people for this sector, and the team that leads our healthcare projects is certified in the course Introduction to SaMD, IEC 62304 and IEC 82304-1, from Medical Device HQ, with 21 hours of continuing education. This article explains what that certificate represents, where the regulatory line falls and what to ask a vendor before signing.

What the certificate in SaMD, IEC 62304 and IEC 82304-1 represents

The three acronyms in the course name answer three different questions, and they are exactly the questions a buyer in this sector needs to ask.

SaMD, Software as a Medical Device, answers "is this software, on its own, a medical device?". It is the concept that defines whether a product falls under the health regulatory regime, and a classification error, in either direction, is the most expensive of all.

IEC 62304 answers "how do you develop and maintain this software in a controlled way?". It is the life cycle standard for medical device software, the one ANVISA cites by name in RDC 657/2022 and the one the FDA accepts as a declaration of conformity in submissions. It organizes planning, requirements, architecture, implementation, verification, release, maintenance, configuration management and problem resolution.

IEC 82304-1 answers "is the complete software product safe for its intended use?". Published in 2016, it covers the safety of health software products that run on general-purpose platforms and go to market without dedicated hardware, spanning the entire cycle, from conception to decommissioning.

The course provider states three objectives: knowing how to create high-quality software, knowing how to create safe software and knowing how to manage software in a way that adds value. In practice, it is the training that allows whoever leads a project to read the instructions for use, classify the software, plan the documentation the standard requires and talk to the manufacturer's quality team in their own language.

It is worth being precise about what this proves. A certificate proves the professional was trained to work within that process. It does not certify the company in any standard and does not certify the product; product conformity is always demonstrated by the manufacturer, with evidence. But it is precisely the vendor that produces that evidence, and producing it well requires someone who knows the standard before the first line of code. Having that knowledge inside the team, and not outsourced to an external consultant, is what separates a vendor that delivers code from a vendor that delivers what goes into the dossier.

Clinic software is not a medical device, and this is where most people get it wrong

The Brazilian medical device market is worth about R$ 75 billion (Brazilian reais) and is among the ten largest in the world, according to a Redirection International survey published by Medicina S/A, a Brazilian healthcare business outlet, in February 2026. A good part of the software this market buys is not a medical device, and confusing the two is expensive in both directions: regularizing what does not need it, or not regularizing what does.

The Brazilian rule is ANVISA's RDC 657/2022 (a resolution of the agency's collegiate board), in force since July 1, 2022. It defines software as a medical device as software that meets the definition of a medical device, intended for one or more medical indications, and that achieves those purposes without being part of a device's hardware. Mobile apps and systems sold by cloud subscription are included if the purpose is medical.

The same resolution states, right in article 1, to whom it does not apply: wellness software; software on the agency's list of non-regulated products; software used exclusively for administrative and financial management in health services; software that processes demographic and epidemiological data without a clinical purpose; and software embedded in a medical device.

The Q&A document ANVISA published in September 2022 goes down to concrete cases. An electronic health record platform that only records and displays information is not a medical device. Scheduling, communication and telemedicine systems, even with images and test results inside the record, are not either. Neither are claims auditing and indicator systems. What changes the answer is a function: when the system starts performing automatic triage, suggesting a diagnosis, suggesting treatment or classifying risk based on guidelines, it becomes SaMD, and ANVISA cites the example of a decision support system that calculates cardiovascular risk and suggests tests and medication, classified as class II or III. The first cut of the classification fits in one question: does the software only record and display, or does it conclude, suggest, calculate or command something with a clinical purpose? What only records and displays stays out.

There is a third group: software that goes into the equipment's own process, because it commands the hardware or because the equipment does not work without it. RDC 751/2022, which now defines the risk classes, is explicit: software whose purpose is to control the hardware of a medical device is not SaMD. It is embedded software or an exclusive-use accessory, and RDC 657 requires it to be regularized together with the equipment, in the manufacturer's process. ANVISA confirms this in the question about software exclusive to an in vitro diagnostic device, installed on an ordinary computer and without which the equipment does not work: it is declared in the equipment's process, not in a separate one, and article 3 of RDC 657 extends the rule to any medical device. Exchanging data with the equipment is a different situation. The same RDC 751 says a SaMD may interact with the hardware of other devices, and ANVISA answered that interfacing software for laboratory equipment is only SaMD when it performs diagnosis or automatic triage, or when it controls the equipment. Software that only receives data from the equipment and, based on it, concludes or suggests something clinical is SaMD in its own right.

For those who buy, this changes who signs what. The one who registers with ANVISA is the legal manufacturer: the company that places the product on the market under its own name, even when the design or manufacturing is done by third parties on its behalf. The vendor that develops for that manufacturer registers nothing; what the manufacturer needs from it is the evidence that goes inside the dossier. The responsibility shifts when the software house itself sells the SaMD as its own product, by license or subscription: in that case it is the manufacturer, it needs an operating authorization, and the requirements of this article become its own.

The same logic applies to what is deliberately kept out of scope. A content portal or an AI assistant aimed at healthcare professionals can be designed not to give diagnoses, treatments, dosages or clinical protocols. This is not generic caution: it is the design decision that keeps the tool on the non-regulated side of the line, and it needs to be made and documented at the start.

Intended use decides the class, and the class decides the entry route

When the software is SaMD, RDC 751/2022 classifies it under Rule 11, which reproduces the structure of the rule with the same number in the European medical device regulation. Software that provides information for diagnostic or therapeutic decisions is class II. If a wrong decision could cause death or irreversible deterioration of health, class IV. If it could cause serious deterioration or surgical intervention, class III. Software that monitors physiological processes is class II, and class III when it monitors vital parameters whose variation puts the patient in immediate danger. Any other SaMD is class I.

The class defines the entry route. Classes I and II go through notification, with a form and a technical dossier kept by the company. Classes III and IV go through registration, with a technical report submitted to the agency. The manufacturer or holder needs an operating authorization in any class and, in classes III and IV, may also need a good manufacturing practices certificate.

Intended use: does the software only record and display, or does it conclude, suggest, calculate or command with a clinical purpose?
Is it part of the hardware? Embedded software or an exclusive-use accessory goes into the equipment's process, with the equipment's class
Impact of error: the severity of what happens if the wrong decision is followed defines the class under Rule 11
Regime: classes I and II by notification; classes III and IV by registration
Four questions, in the order RDC 657/2022 and RDC 751/2022 ask them. The answer to the first usually ends the conversation for health records, scheduling and portals.

A hospital that develops its own software has a separate regime. Article 5 of RDC 657 exempts from regularization SaMD developed in-house, for the exclusive use of the health service, in classes I and II, as long as it does not interfere with a medical device subject to regularization. The trade-off is documentary: complete validation records, with change history, kept for at least ten years after decommissioning, or else the software is considered non-regularized. For a hospital that hires a company to build that "in-house" software, the consequence is direct: being exempt from registration does not exempt it from evidence, and the hospital is the one accountable for it.

What ANVISA asks for in the dossier, and what that means for those who buy

Article 12 of RDC 657 lists what the technical report of a class III or IV SaMD needs to include beyond what applies to any device: software architecture; hardware architecture and minimum technical requirements; platform; compatibility, interoperability and communication with other products; cybersecurity architecture and controls; verification and validation; risk management; residual anomalies identified and how they are mitigated; clinical evaluation, with a description of the algorithms; and the declaration of conformity with international standards.

Article 13 says which standards, at a minimum, that declaration covers: IEC 62304, on software life cycle; IEC 62366-1, on usability; and ISO 14971, on risk management, in the versions cited in the resolution or more recent versions. ANVISA later clarified that the declaration is required for classes II, III and IV, that it must be a document signed by the manufacturer, and that a simple list of standards used in the dossier does not count as a declaration. Whoever does not declare one of the standards must justify it and deliver, in its place, the document corresponding to the missing standard: the product life cycle description for IEC 62304, the usability report for IEC 62366-1 and the risk management report for ISO 14971.

There are also requirements on what accompanies the product. Article 7 requires the instructions for use, or the software itself, to include update procedures, minimum hardware and software requirements, the operating principle with a description of the algorithms, alerts and warnings, interoperability specifications and cybersecurity information. Article 8 asks for product and version identification that allows production to be traced, in place of a batch number. And article 16 reaches into the product's life after registration: a change that creates new functionality or a new clinical indication, that significantly affects safety or efficacy, or that alters the software's visual identity must be petitioned to the agency. For those who buy, the vendor's release plan needs to separate, from the start, what is maintenance from what is a change that goes through ANVISA.

And RDC 848/2024, which sets the essential safety and performance requirements, closes the circle in article 50: the software of a medical device, or software that is itself the device, "must be developed, manufactured and maintained in accordance with the state of the art, taking into account the principles of development life cycle, risk management, including information security, verification and validation". The following articles ask for minimum network and IT security requirements declared by the manufacturer and protection against unauthorized access.

Read that list through a buyer's eyes. None of its items says "the vendor must have certificate X". All of them say "the manufacturer must demonstrate how the software was made". The one who signs the declaration is the manufacturer. The one who produces most of the evidence, when development is outsourced, is whoever develops. If the vendor delivers the working system and does not deliver the architecture, traceable requirements, test protocol and report, and the list of known anomalies, the manufacturer has nothing to sign with. That is why the training of whoever coordinates the project matters: it is what makes that evidence come into being alongside the code, rather than being reconstructed afterward.

IEC 62304 translated: three safety classes and what each requires

The central IEC 62304 concept for buyers is the software safety class, which is not the same as the device's risk class at ANVISA. The standard has three: A, when the software cannot contribute to a hazardous situation, or when it can contribute but the risk is acceptable after verified risk control measures external to the software; B, when it can contribute even after those measures, but the possible harm is not serious; and C, when the possible harm is serious or can lead to death. The class defines how much documentation and verification the standard requires.

What IEC 62304 requiresClass AClass BClass C
Documented software requirementsYesYesYes
Software architectureNoYesYes
Detailed unit designNoNoYes
Unit verificationNoYesYes
System testingYes, since the 2015 amendmentYesYes

Source: IEC 62304:2006 with amendment 1 of 2015, title and amendment checked on the official IEC page; requirements by class according to the Johner Institute's summary of safety classes, read on October 2, 2026. Risk control measures external to the software can lower the class, as long as they are verified.

In practice, the class is a joint decision with the manufacturer's quality team, made before the code, and the manufacturer's internal quality procedure may ask for more than the standard requires for that class. When in doubt, the governing document is that procedure, not the vendor's process preference. That is how it works when the vendor enters the process of whoever registers, and it is the kind of conversation IEC 62304 training prepares you for.

Two neighboring concepts come right after. The first is SOUP, the standard's acronym for software of unknown provenance: every library, framework and third-party component that goes into the product. The standard requires identifying the name, version and manufacturer of each one, assessing known anomalies and justifying why that component is fit for the intended use. For buyers, this means choosing a framework stops being a developer's preference and becomes a documented decision. A vendor that cannot list what it uses cannot build the component inventory that US law now requires for connected devices.

The second is the division of labor between sister standards. IEC 82304-1 covers the health software product as a whole, including installation, maintenance and decommissioning. IEC 81001-5-1, from 2021, covers cybersecurity activities throughout the life cycle: threat modeling, secure development, vulnerability handling and patch management. ISO 14971 covers risk and IEC 62366-1 covers usability, and 62304 refers to both.

The documentation a manufacturer expects from its developer

When a medical device manufacturer outsources software, the list of what it expects to receive follows the structure of the standards themselves. It is the list buyers should put in the proposal, not discover at delivery.

StandardWhat the manufacturer expects to receive
IEC 62304Life cycle: development plan, architecture, requirements, traceability, test specifications and verification reports
ISO 14971Risk: software-related risk documentation, aligned with the device's risk analysis
IEC 81001-5-1Cybersecurity: threat model, secure development documentation, vulnerability handling and patch management
Data protectionDocumentation of compliance with each market's personal data legislation, such as LGPD, GDPR and HIPAA
IEC 82304-1Product: health software product documentation, including installation, maintenance and decommissioning
IEC 62366-1Usability: usability engineering documentation, including intended use and foreseeable use errors

Documentation families organized by the standards RDC 657/2022 cites and by the sister product and cybersecurity standards, with our description of the content of each.

Four practical consequences come out of this list.

Traceability becomes the backbone of the project. User needs become test cases, and each feature gets its entire path written down: need, requirement, architecture decision, code, test and result. Tests and documentation need to live where the manufacturer's quality team can always get the latest version, without asking.

The instructions for use become a vendor deliverable. It is the kind of document an ordinary software vendor never writes and that, in this sector, article 7 of RDC 657 requires, cybersecurity section included.

Data protection changes the API. A HIPAA, GDPR or LGPD (Brazil's data protection law) review usually makes a patient listing return only the minimum fields needed to draw the screen, instead of the entire record. It is a small engineering decision that only shows up when someone looks at the product through the lens of whoever will be audited.

And infrastructure requirements arrive in writing. Backup, in an ordinary project, is an operations item. In a regulated one, it is a requirement with an owner, a test and a record.

For the buyer, the conclusion is simple. If the vendor delivers the repository link and a video of the screen working, you bought code. If it delivers the traceability chain, the component risk analysis, the third-party component inventory, the test protocols and reports and the instructions for use, you bought what goes into your dossier. The price difference between the two deliveries exists, and it is smaller than the cost of reconstructing the evidence later, when the auditor asks.

If the product is going to the United States

A Brazilian manufacturer that exports, or a foreign manufacturer that hires development in Brazil, finds at the FDA the same logic under different names.

The agency adopts the SaMD definition from the IMDRF, the international forum of regulators, and it says the same as RDC 657's, so the classification done for ANVISA serves as a starting point. The difference that matters for buyers lies in the documentation. The guidance on the content of software submissions, published in June 2023, works with two levels. The enhanced level applies when a software failure could create a hazardous situation with a probable risk of death or serious injury, assessed before risk control measures; the basic level applies to everything else, with two exceptions the guidance itself lists: blood establishment software gets the enhanced level in any case, and so do class III devices under the US classification, as well as the device part of combination products with a drug or biologic, unless there is a detailed justification in the submission. At both levels, the FDA expects a software description, a risk management file, a requirements specification with traceability, architecture, verification and validation tests, version history and the list of unresolved anomalies with an impact assessment. And it accepts, in place of the summary of development practices, a declaration of conformity with IEC 62304.

Two recent changes increase the weight of this. Since February 2, 2026, the FDA's new quality system regulation, the QMSR, has been in force, incorporating ISO 13485:2016 by reference. And section 524B of US law, for so-called cyber devices, requires in submissions a plan to monitor and address vulnerabilities after launch, processes that provide reasonable assurance of security, and an inventory of software components, the SBOM. The FDA's cybersecurity guidance was updated in February 2026.

For buyers, the consequence is about schedule and cost. A vendor that maintains the component inventory and traceability from the first sprint delivers the US submission documentation as a byproduct of the work. A vendor that does not turns the submission into a reconstruction, months later, done by people who did not write the code.

What to ask the vendor before signing

The manufacturer does not hire by badge. It hires by the evidence it can put into its dossier, and the best way to know whether it will receive it is to ask early: which IEC 62304 safety class do you propose for this component, and why; how do you record third-party components and known vulnerabilities; where will traceability live and who will have access to it; and who on the team is trained in the standards ANVISA cites. These are four questions any buyer in this sector can ask in the proposal, and they separate those who already know a manufacturer's quality process from those who only make promises.

At X-Apps, these questions have answers before the contract. With people trained and certified in SaMD, IEC 62304 and IEC 82304-1 leading our healthcare projects, we build the documentation plan with the client's quality team before the first line of code and deliver the evidence together with the software. The process does not guarantee approval by any agency. What it does is have the answer written down before the auditor asks: how the software was specified, what risk it introduces and how it was verified.

Sources

  • ANVISA, Resolution RDC No. 657, of March 24, 2022, consolidated text, read on October 2, 2026; and Questions and Answers on RDC 657/2022, version 1, of September 1, 2022.
  • ANVISA, Resolution RDC No. 751, of September 15, 2022, definitions and Rule 11 of Annex I; and Resolution RDC No. 848, of March 6, 2024, articles 49 to 53.
  • IEC, official pages for the publications IEC 62304:2006+AMD1:2015, IEC 82304-1:2016 and IEC 81001-5-1:2021, read on October 2, 2026; summary of safety classes by the Johner Institute.
  • FDA, Content of Premarket Submissions for Device Software Functions, final guidance of June 2023; Quality Management System Regulation page; Cybersecurity in Medical Devices guidance page, February 2026 version. Read on October 2, 2026.
  • Medicina S/A, February 19, 2026, Redirection International analysis of the size and position of the Brazilian medical device market.
  • Medical Device HQ, course page for Introduction to SaMD, IEC 62304 and IEC 82304-1.

Frequently asked questions

No, as long as it only records and displays information. ANVISA (Brazil's health regulator) answered this in the Q&A on RDC 657/2022: health records, scheduling and telemedicine, even with images and test results, are not medical devices. The system becomes SaMD when it includes automatic triage, diagnosis suggestions or treatment suggestions.

Previous post
AI that sorts tickets costs pennies: what else goes into the bill
Next post
Your site may be out of ChatGPT search without anyone deciding it
Newsletter

One email a month, no noise

What we learn shipping custom software and applied AI.

Related articles

How to turn your company into an exponential organization3 min · Business
What to consider when validating a software development project2 min · Business
Open and closed scope: software and the urgency of innovation1 min · Business
Basic API guide for companies3 min · Business
The power of collective intelligence in software development1 min · Business

Accelerate your business with X-Apps

Ask for a Budget
X-Apps is an IT provider partner and advised by
Gartner
Receive our e-mails
News on software and applied AI.
Follow X-Apps